GEOMAR moorings are typically equipped with instruments recording pressure, temperature, conductivity, dissolved oxygen and current velocity. Instruments with pressure, temperature, conductivity and oxygen sensors were calibrated in situ immediately prior to and after a mooring deployment period by attaching them to the CTD frame during CTDO casts. Correction terms were then developed from the difference between the sensor readings and the calibrated CTDO data during several minute long calibration stops. These correction terms were then applied to the full deployment periods. This ensured best data quality with recognition of potential sensor drifts and also allowed for the estimation of calibration and measurement errors (Hahn et al. 2014, Bittig et al. 2018, Berx et al. 2019). Moored Acoustic Doppler Current Profiler bin depths were corrected using the sound speed at instrument depth following the approach by Shcherbina et al. (2005). Velocities were not corrected, but respective measurement errors were assumed as described in Hahn et al. (2014). For all instruments within a mooring that did not record pressure, full deployment pressure time series were estimated by linearly interpolating between the instruments having a pressure sensor.
instrument = Label:kpo_1140_01 File:kpo_1140_instr_01.nc Type:Teledyne RDI QuarterMaster 150 SN:14911instrument = Label:kpo_1140_02 File:kpo_1140_instr_02.nc Type:GEOMAR TD-Logger SN:68instrument = Label:kpo_1140_03 File:kpo_1140_instr_03.nc Type:Teledyne RDI LongRanger 75 SN:02627instrument = Label:kpo_1140_04 File:kpo_1140_instr_04.nc Type:Seabird MicroCat 37-SM SN:02248 depth/pressure data was at least partially derived and not directly measured, consult quality flags for detailsinstrument = Label:kpo_1140_05 File:kpo_1140_instr_05.nc Type:Aanderaa Optode 3830 SN:00938 depth/pressure data was at least partially derived and not directly measured, consult quality flags for detailsinstrument = Label:kpo_1140_06 File:kpo_1140_instr_06.nc Type:Seabird MicroCat 37-IM mit P SN:02717instrument = Label:kpo_1140_07 File:kpo_1140_instr_07.nc Type:Aanderaa Optode 3830 SN:01140 depth/pressure data was at least partially derived and not directly measured, consult quality flags for detailsinstrument = Label:kpo_1140_08 File:kpo_1140_instr_08.nc Type:Sontek Argonaut SN:187 depth/pressure data was at least partially derived and not directly measured, consult quality flags for detailsinstrument = Label:kpo_1140_09 File:kpo_1140_instr_09.nc Type:Nortek Aquadopp Seriell SN:P26209-34instrument = Label:kpo_1140_10 File:kpo_1140_instr_10.nc Type:Aanderaa RCM 11 SN:00477 depth/pressure data was at least partially derived and not directly measured, consult quality flags for detailsinstrument = Label:kpo_1140_11 File:kpo_1140_instr_11.nc Type:McLane Profiler MMP SN:11617Quality flags:0:unknown1:good_data2:probably_good_data3:potentially_correctable_bad_data4:bad_data5:not_used6:not_used7:nominal_value8:interpolated_value